Chapter 4: Test and Calibration Procedures ABL77 Service Manual
Testing the Valve Connections, Continued
NOTE:
(continued)
Step Action
9.
Remove the index card, or equivalent, from the door slot and the soft
cloth or gauze from the manifold ports.
10.
The calibration resistors may have been inadvertently affected during
this procedure. To verify proper calibration settings, Refer to the
procedure ā€œCalibrating the SC and Hct Circuitsā€.
11.
Shut down and re-assemble the analyzer.
4-12 Rev. F