ABL77 Service Manual Chapter 2: Analyzer Description
Calibration
Introduction
The user can define the frequency of automatic two-point calibrations. The
possible intervals are 1, 2, 3 or 4 hours. The Calibration option on the Main
Menu is used to initiate a manual two- point calibration as desired.
Hydration
If the calibration is more than 30 minutes past the scheduled time, a preliminary
flush of Cal1 solution is performed prior to the two-point calibration. A message
“hydrating….” will be displayed.