Chapter 7: Re-Assembly ABL77 Service Manual
Backlight Inverter Board
Introduction
Follow these steps to re-assemble the backlight inverter board to the upper module
as shown in Figure 7-3.
Step Action
1.
Hold the backlight inverter board with the single center connector
facing the left.
2.
Re-connect the LCD display cables to the board at the matching ends
with the black wires facing away from you.
3.
Attach the board to the upper module with two screws.
4.
Apply insulating tape across the length of the backlight inverter board
to prevent electrical shock.
Figure 7-3
Part Function
1. Backlight inverter cable
2. Backlight inverter board screws
3. Insulation tape
4. Backlight inverter board
5. LCD display cables