ABL77 Service Manual Chapter 4: Test and Calibration Procedures
Valve Test Procedures
Introduction
The following procedures are used to determine which valve is malfunctioning on
a valve board with replaceable valves.
Evaluation
Use the following diagrams to determine which type of fluidics system is installed
in the analyzer.
Remove the lower module and compare the tubing arrangements to the analyzer
being serviced, then proceed to the valve test procedure for the appropriate fluidics
system.
Continued on next page