ABL77 Service Manual Chapter 1: Introduction
Test Equipment and Tools
Required
equipment and
tools
To carry out the procedures for servicing the analyzer, the following test
equipment and tools are required: Parts marked with “N/A” in the REF column are
to be purchased locally.
REF Description Use
902-6XX SCi sensor cassette Testing analyzer after repair
944-069 Calibration solutions pack (cal pack) Testing analyzer after repair
902-578 Electronic zeroing fixture Zeroing sensor channels
913-741 Manager setup disk Custom setup of system features
920-721
5
/
64
” ball point Allen wrench Various applications
920-722
1
/
16
” ball point Allen wrench Lower module
920-723
7
/
64
” ball point Allen wrench Lower module
920-726
5
/
16
” open-end wrench Electronics module
920-727
5
/
16
” nut driver Electronics and Lower modules
920-728 ¼” nut driver Electronics module
920-729 Valve board calibration cassette Calibrating the Hct and SC circuit
991-260 Loctite 401 Adhesive Securing the handle to the housing
991-264 Silicone adhesive, RTV medical grade Securing cables to prevent disconnect
991-265 Loctite 222 Threadlocker Securing assorted screws and nuts
905-674 Sample path obstruction tool kit Various applications
943-906 Cleaning solution Cleaning the waste drain
N/A Deionized water (or sterile water) Various applications
N/A Soft cloth, gauze or other absorbent material Various applications
N/A Protective attire (gloves, safety glasses, and lab
coat)
Various applications
N/A Flat head screwdriver
Lower module
N/A Small flat head screwdriver Adjusting the calibration settings
N/A Phillips head screwdriver Various applications
N/A Needle nose pliers Lower module
N/A Scissors or equivalent trimming tool Various applications
N/A AT style keyboard Entering system commands
N/A Digital multimeter Testing electronics
N/A Digital voltmeter (or multimeter) 5 volt power output setting
N/A 3½” formatted diskettes Downloading data files
N/A Index card or equivalent Block the opto sensor
Rev. F 1-5