ABL77 Service Manual Chapter 8: Maintenance
Long-Term Storage
Introduction
This topic describes the necessary steps to flush an analyzer before any period (>24
hours) of long-term storage.
CAUTION: Pushing the syringe plunger too hard can damage internal tubing
connections and cause leaks.
WARNING/
CAUTION:
CAUTION: Be prepared to collect waste fluids flowing from open luers.
NOTE:
Salt crystals may form periodically in the analyzer fluid path if the analyzer is
turned off for extended periods of time. This tendency is accelerated if a cal pack
and sensor cassette are not left on the analyzer. These crystals will form
obstructions in the fluid path.
Long-term
storage
The following steps should be performed prior to long-term storage of the ABL77.
Step Action
1.
From the Main Menu press Options.
Enter a Password as necessary.
2.
At the Options menu select Fluidics.
RESULT: The following screen appears:
Figure 8-6
Select Long-Term Storage.
Press Yes to proceed.
Continued on next page
Rev. F 8-9

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL 77 Series and is the answer not in the manual?

RADIOMETER ABL 77 Series Specifications

General IconGeneral
BrandRADIOMETER
ModelABL 77 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals