Chapter 8: Maintenance ABL77 Service Manual
Long-Term Storage, Continued
Long-term
storage
(continued)
Step Action
3.
If a sensor cassette is… then…
attached the analyzer will remove any fluid then
prompt you to remove the sensor
cassette. The analyzer will sense when
the cassette has been removed and will
automatically proceed to the next
screen.
not
attached the analyzer will automatically proceed
to the next screen.
4.
There are three components of the ABL77 that become contaminated
with sample residue with use. These components are the sensor
cassette, the waste line and the waste pouch within the cal pack. The
sensor cassette and cal pack can be discarded as biohazardous waste.
5.
To decontaminate the waste line ensure there is a cal pack in place to
collect waste fluids. Introduce 1 cc of Cleaning Solution (REF: 943-
906) into the waste drain (the black opening located directly below the
inlet probe when a cassette is in place).
6.
Activate the waste pump by pressing Pump to carry this fluid
through the waste line and into the cal pack.
Press Continue.
7.
Introduce 1 cc of deionized water into the waste drain.
Press Pump to clear the fluid.
Press Continue.
8.
Remove the cal pack
Press Continue.
Continued on next page
8-10 Rev. F

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL 77 Series and is the answer not in the manual?

RADIOMETER ABL 77 Series Specifications

General IconGeneral
BrandRADIOMETER
ModelABL 77 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals