ABL77 Service Manual Chapter 5: Dismantling
Waste Pump Head and Waste Pump, Continued
Removing the
waste pump
Follow these steps to remove the waste pump from the back of the lower module as
shown in Figure 5-16.
Step Action
1.
Remove the lower module.
2.
From the backside of the lower module, stand the two flanges straight,
where the waste pump cable is connected to the motor. This will allow
the motor to slide out the front of the module.
3.
Disconnect the cable connector.
4.
From the front of the lower module remove the waste pump head.
5.
Using a Phillips head screwdriver, remove the two mounting screws
from the pump motor in the front of the lower module.
6.
Pull the waste pump forward.
Continued on next page
Rev. F 5-23