EasyManua.ls Logo

RADIOMETER ABL 77 Series - Page 98

RADIOMETER ABL 77 Series
355 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4: Test and Calibration Procedures ABL77 Service Manual
Calibration Lines Procedure, Continued
Calibration lines
procedure
(continued)
Step Action
17.
Press Continue.
RESULT: The following screen will appear.
18.
Refill the syringe with deionized or sterile water and flush the Cal2
line in a similar manner, pressing Cal2 to open a pathway between
the cassette luer and the Cal2 manifold luer.
Repeat as necessary.
Press OK when flushing is complete.
19.
Re-attach the cal pack and sensor cassette (if one is currently
installed).
Press OK.
RESULT: The system will return to the Options menu.
20.
Perform a manual calibration to ensure proper function prior to sample
analysis.
4-10 Rev. F

Table of Contents

Related product manuals