241
mass mass value in current unit is returned,
tare tare value in calibration/adjustment unit is returned,
unitdetermines current (displayed) measuring unit:
Byte no.
Unit
B5 B4 B3 B2 B1
B0 Dec
g
gram
0
0
0
0
1
0
1
kg
kilogram
0
0
0
0
0
1
2
ct
carat
0
0
0
1
0
0
4
lb
pound
0
0
1
0
0
0
8
oz
ounce
0
1
0
0
0
0
16
N
Newton
1
0
0
0
0
0
32
For instance:
Byte no B5 B4 B3 B2 B1 B0
Value 0 0 0 0 1 0
Scale use kilogram [kg].
status determines scale’s status
Status
bytes
Task Dec
0
correct measurement (no error message from scale)
1
1
stable measurement
2
2
scale in zero indication
4
3
scale tarred
8
4
scale in second measuring range
16
5
scale in third measuring range
32
6
scale reports NULL error
64
7
scale reports LH error
128
8
scale reports FULL error
256
For instance:
byte no. B8 B7 B6 B5 B4 B3 B2 B1 B0
value
0 0 0 0 1 0 0 1 1