cobas p 512 Part B - 0BSystem description
Operator's Manual - Version 1.6 - 10/2015 2-73
Material no.
The material ID as configured for the identified tube type.
Quantity
Sample volume in the tube in µl. The sample volume is calculated by the system based on fill
level, sample liquid color and tube dimensions.
Class
This field gives a quality statement on the detected tube and its sample content, for example
“Serum ok”. The camera system is calibrated to assess the sample quality (only available with
the QS I module).
The value returned from the camera module differentiates between the following classes:
returned value
displayed class
0
Serum ok
1
hemolytic
2
lipaemic
3
icteric
4
unknown
Return code of QS I
In case the camera's classification of the serum has failed, a return code is sent representing
the error that has occurred.
For a list of all return codes refer to chapter 'Troubleshooting'.
Number of turns
The number displayed here refers to the number of turn movements the lifting gripper had
to perform before the tube was correctly identified.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas p 512 and is the answer not in the manual?

Roche cobas p 512 Specifications

General IconGeneral
BrandRoche
Modelcobas p 512
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals