Part E - Troubleshooting cobas p 512
5-16 Operator's Manual - Version 1.6 - 10/2015
Error standstill
monitoring
Error standstill monitoring for
positioning commands. Axis
deflection at standstill in excess
of the tolerance. The buffer
does not change when this
error occurs. Causes: bad rule
parameters, axis was deflected
Press the [Repeat] button.
Contact a service technician if the error
persists.
No liquid level
found
Motor control board awaited
an interruption signal from the
liquid level board/cap
recognition. No interruption
signal was entered. The entire
Check liquid level position (height, start of
liquid level ...)
Error in liquid level
Error in liquid level detection.
The entire buffer is deleted
Check liquid level sensors, possibly also for
too little material.
A Transfer error
Communication error from the
computer to the DMS/SMS
control boards.
Red fail LED is blinking (%1).
Check the cabling to the motor control
board.
Press the [Repeat] button.
Contact a service technician if the error
Check sum error in
control telegram
Checksum error in control
telegram.
Communication error from the
computer to the DMS/SMS
control boards.
Red fail LED is blinking (%1).
Check the RS485 interface (hardware).
Press the [Repeat] button.
Contact a service technician if the error
persists.
Error command
format - control
Command format error in the
control telegram. The entire
buffer is deleted. (%1)
Check the interface protocol.
Press the [Repeat] button.
Contact a service technician if the error
Check sum error in
the instruction
telegram
instruction telegram.
Communication error from the
computer to the DMS/SMS
control boards.
Red fail LED is blinking (%1).
Check the RS485 interface (hardware).
Press the [Repeat] button.
Contact a service technician if the error
persists.
Error command
format - instruction
Command format error in the
instruction telegram. The
buffer is not changed. (%1)
Check the interface protocol.
Press the [Repeat] button.
Contact a service technician if the error