Part E - Troubleshooting cobas p 512
5-18 Operator's Manual - Version 1.6 - 10/2015
Error message
Description
Remedy
078
L Initialization
error
(L) Initialization error.
During initialization, an axis'
limit switch hasn't been
reached or the monitored time
limits have been exceeded.
The entire buffer is deleted
when this error occurs (%1).
If this error occurs repeatedly, contact a
service technician: Make sure the axes are
running smoothly. Check belt tension (if
applicable for the affected axis). Check the
limit switch. Check the voltage supply (5V,
24V).
079
Limit switch error
(moved into limit
switch)
Error and limit switch:
The limit switch has been
triggered or has not been
deactivated.
Any active positioning is
terminated and the entire
buffer deleted whenever this
error occurs (%1).
Check the drive system and limit switch.
Press the [Repeat] button.
Contact a service technician if the error
persists.
080
Optical waveguide
interrupted
The optical waveguide has been
interrupted. You can no longer
search for a cap. (%1)
Contact a service technician.
081
Data buffer
overflow
Data buffer overflow!!!
You may not be able to
continue working without
errors!!!
The entire buffer is deleted
when this error occurs. (%1)
Reboot the system.
Contact a service technician if the error
persists.
082
Defined maximum
position exceeded
The programmed position
exceeds the maximum position
defined in the motor
parameters. (%1)
Use the log file to determine which
position the system should be moved to. If
this position is correct, adjust the
maximum position for the motor
parameter set. Contact a service technician
if the error persists.
083
Error PID filter
Error at motor control board
DMS (Address=%1):
PID filter error in the internal
controller.
Bad value in the regulation
parameters.
Press the [Repeat] button.
Contact a service technician if the error
persists.
Check the regulation parameters. Replace
DMS control board.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas p 512 and is the answer not in the manual?

Roche cobas p 512 Specifications

General IconGeneral
BrandRoche
Modelcobas p 512
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals