Part E - Troubleshooting cobas p 512
5-18 Operator's Manual - Version 1.6 - 10/2015
L Initialization
error
(L) Initialization error.
During initialization, an axis'
limit switch hasn't been
reached or the monitored time
limits have been exceeded.
The entire buffer is deleted
when this error occurs (%1).
If this error occurs repeatedly, contact a
service technician: Make sure the axes are
running smoothly. Check belt tension (if
applicable for the affected axis). Check the
limit switch. Check the voltage supply (5V,
24V).
Limit switch error
(moved into limit
switch)
The limit switch has been
triggered or has not been
deactivated.
Any active positioning is
terminated and the entire
buffer deleted whenever this
Check the drive system and limit switch.
Press the [Repeat] button.
Contact a service technician if the error
persists.
Optical waveguide
The optical waveguide has been
interrupted. You can no longer
Contact a service technician.
Data buffer
overflow
You may not be able to
continue working without
errors!!!
The entire buffer is deleted
when this error occurs. (%1)
Contact a service technician if the error
persists.
Defined maximum
position exceeded
exceeds the maximum position
defined in the motor
parameters. (%1)
Use the log file to determine which
position the system should be moved to. If
this position is correct, adjust the
maximum position for the motor
parameter set. Contact a service technician
Error PID filter
Error at motor control board
DMS (Address=%1):
PID filter error in the internal
controller.
Bad value in the regulation
Press the [Repeat] button.
Contact a service technician if the error
persists.
Check the regulation parameters. Replace
DMS control board.