cobas p 512 Part E - Troubleshooting
Operator's Manual - Version 1.6 - 10/2015 5-49
ISAM: Record '%'
locked
ISAM: End of file or
range
You have reached the last (first)
data record in the general
parameters programs and
Select [Back] ([Next]) if you have reached
the last (first) data record.
ISAM: Duplicate
key within primary
ISAM: No rights for
ISAM: File '%' not
open
You can only read but not open
this file.
This is a system error. If restarting the
program does not cure this problem, check
on the operating system level for possible
Record copied to
<%> !
copying coordinate data
records from and to the
Deleting data has
You cannot delete this data
record!
Transfer
parameters of a
program are
incorrect
A transfer parameter error
occurred: Required parameter
was not stated. Incorrect
transfer type. Value/contents
invalid.
In general the programs are started via
general parameters of background and
program parameters. Use the log file to
determine appropriate records and check
the transfer parameters. For the valid
transfer parameters, refer to 'use <program
name>' or the software documentation.
Bad entry
Operator has made a bad entry
or does not have access to the
option that is not supported by
the program.
Check if the program supports all the
menu items or input forms that have been
cleared for the menu file or the screen file,
respectively. Check the entry in the log file
on troubleshooting. Contact a service