Remote control commands
R&S
®
SGS100A
241User Manual 1173.9105.02 ─ 13
Return values:
<TotalPhysMemKb> integer
Total physical memory.
<ApplicMemKb> integer
Application memory.
<HeapUsedKb> integer
Used heap memory.
<HeapAvailableKb> integer
Available heap memory.
Usage: Query only
:MMEMory:LOAD:STATe <SavRclStateNumb>, <file_name>
Loads the specified file stored under the specified name in an internal memory.
After the file has been loaded, the instrument setting must be activated using an *RCL
command.
Setting parameters:
<SavRclStateNumb> Determines to the specific <number> to be used with the *RCL
command, e.g. *RCL 4.
<file_name> String parameter to specify the file name with extension
*.savrcltxt.
Example: See "Storing and Loading Current Settings" on page 236.
Usage: Setting only
:MMEMory:MDIRectory <Directory>
Creates a subdirectory for mass memory storage in the specified directory. If no direc-
tory is specified, a subdirectory is created in the default directory. This command can
also be used to create a directory tree.
Setting parameters:
<Directory> string
String parameter to specify the new directory.
Example: See "Working with Files and Directories" on page 236.
Usage: Event
:MMEMory:MOVE <SourceFile>, <DestinationFile>
Moves an existing file to a new location or, if no path is specified, renames an existing
file.
MMEMory subsystem