Remote control commands
R&S
®
SGS100A
264User Manual 1173.9105.02 ─ 13
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
The command fills the selected user correction list with the level values measured by
the power sensor for the given frequencies.
To select the used power sensor set the suffix in key word SENSe.
Example:
CORR:CSET:DATA:SENS:POW:SONC
fills the user correction list with level values acquired by the
power sensor connector to the [SENSOR] connector.
Usage: Event
Manual operation: See "Fill User Correction Data with Sensor" on page 112
[:SOURce]:CORRection:CSET:DELete <Filename>
Deletes the specified table.
The lists are stored with the fixed file extensions *.uco in a directory of the user's
choice. The directory applicable to the commands is defined with the command
MMEMory:CDIR. A path can also be specified in command SOUR:CORR:CSET:CAT?,
in which case the file in the specified directory is deleted.
Setting parameters:
<Filename> <table name>
Example:
MMEM:CDIR '/var/sgs/ucor'
selects the directory for the user correction files.
CORR:CSET:DEL 'UCOR1'
deletes the table ucor1.
Usage: Setting only
Manual operation: See "User Cor. Data" on page 106
[:SOURce<hw>]:CORRection:DEXChange:AFILe:CATalog?
Requests a list of available ASCII files for export/import of user correction data. The
individual files are separated by commas.
The ASCII files are stored with the fixed file extensions *.txt or *.csv in a directory
of the user's choice. The directory applicable to the commands is defined with the com-
mand MMEMory:CDIR.
Return values:
<Catalog> string
SOURce:CORRection subsystem

Table of Contents