8015416/ZPN2/2017-09-08 • © SICK AG • Subject to change without notice 69
Operating instructions Distance measuring device DL100 – CANopen®
Operation via Ethernet (Ethernet interface)
Field "Advanced device functions"
Parameter Description
Averagelter
distance
Selectlterdepthforthedistancevalues.
Read/Write access
Read and write
Input value
0: fast / 1: medium / 2: slow
Factory setting
Medium
Note: Medium and Fast use the same averaging depth.
Averagelter
velocity
Selectlterdepthforthespeedvalues.
Read/Write access
Read and write
Input value
0: fast / 1: medium / 2: slow
Factory setting
Medium
Error
rejection
Select time for error suppression. If there is an error, the
measurement value is indicated as "0".
Read/Write access
Read and write
Input value
0:o/1:50ms/2:200ms
Factory setting
200 ms
Table 42: Page “Parameter settings” – eld “Advanced device function”
Field "Heater" Requirements for the display
Devices with the option "Heating" (DL100-xxHxxxxx)
Parameter Description
Heater thresh-
old
Enter power up threshold for heating.
Read/Write access
Read and write
Input value
–10...+40
Unit
°C
Factory setting
–10
Table 43: Page "Parameter settings" – eld "Heater"