ACUSON Cypress™ Operators Manual 6-13
Using the Cardiac Calculations Package
Making Corrections to the Trace and Area Measurements
No matter which method you choose to use to perform the measurements, cor-
recting the Trace or Area outline is the same. Press the <
BACKSPACE> key to
erase a small portion of the outline. Press and hold the <
BACKSPACE> key to
continuously erase the path of the cursor until the <
BACKSPACE> key is
released.
Stopping a Measurement Before Completion
To stop a measurement or sequence of measurements and return the cursor
to the currently displayed Calculation menu, press the <
ESC> key before
completing the measurement.
To stop a measurement or sequence of measurements and exit the Calcula-
tion package, press the
CALC key before completing the measurement.
Setting the Trace Resolution
The Resolution feature defines the spacing of the dots used to form the trace
outlines. To change the resolution, follow these steps:
1. Press SETUP, double-click the Calculations heading, and highlight Mea-
surements.
2. Specify resolution settings in the 2D Area and Doppler Trace boxes. Using
the Resolution pull-down menu, select a number from 1 to 8. A lower number
specifies a more vivid resolution (i.e., more dots); a higher number specifies
a less vivid resolution.
Viewing the Patient Report
There are two ways of viewing the Patient Report:
While in the calculation package, select Report from the calculation Selection
Menu.
At any time, press the PATIENT function key and click the Reports tab.