ACUSON Cypress™ Operators Manual 4-21
Using Existing Studies
Batch Removal
To remove a group of patients or studies, follow these steps:
1. Press the PATIENT function key to display the Patient/Study list.
2. To highlight a group of studies in consecutive order, use the <SHIFT> key.
3. To highlight studies that are not in consecutive order, use the <CTRL> key.
4. Click the [REMOVE] button to delete the studies from the system hard drive.
5. A system message prompts you to specify what data should be removed by
clicking the corresponding radio button: All Data (deletes all information per-
taining to the patient), All Images (deletes all loops and frames; the patient
name and patient information remain in the Patient/Study List), or Unflagged
Images (all image data not flagged will be deleted).
6. Click [YES TO ALL] to remove all of the highlighted studies or click [YES] to
confirm deletion of each study, one-at-a-time.
7. To exit, press 2D or any Mode key except COLOR.
USE THE <SHIFT>
KEY ON THE
CYPRESS KEYPAD
TO HIGHLIGHT MUL-
TIPLE STUDIES. THE
CYPRESS SYSTEM
WILL REMOVE ALL
THE HIGHLIGHTED
STUDIES.
USE THE <CTRL>
(CONTROL) KEY TO
HIGHLIGHT MULTI-
PLE STUDIES THAT
ARE NOT IN CON-
SECUTIVE ORDER.
THE CYPRESS WILL
REMOVE ONLY THE
HIGHLIGHTED
STUDIES.