Transferring Studies
9-26 ACUSON Cypress™ Operator’s Manual
Troubleshooting an MO Disk Transfer
When transferring studies to an MO disk, an error message may be displayed on the
Shutdown page. A red light appears in the upper right corner of the screen indicating an
error has occurred.
Table 46 describes the possible errors and solutions.
TABLE 46. MO Disk Errors
Error Solution
No disk in the system
Insert a disk.
The disk is write-protected
Insert a disk that is not write-pro-
tected.
Insufficient space on the disk
for the study
Insert a disk with more available
space or delete studies from this
disk.
A disk with an unknown format
is detected
Format this disk or insert a properly
formatted disk.
An unformatted disk is in the
drive
Format this disk or insert a properly
formatted disk.
An MO disk containing studies
from a source other than a
Cypress has been inserted. You
will not be able to save stud-
ies to this disk.
If the studies on this disk are no
longer needed, format the disk from
the utilities menu and then save the
study. Otherwise, insert a disk con-
taining only Cypress studies or a
new MO disk.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ACUSON Cypress and is the answer not in the manual?

Siemens ACUSON Cypress Specifications

General IconGeneral
BrandSiemens
ModelACUSON Cypress
CategoryMedical Equipment
LanguageEnglish

Related product manuals