ACUSON Cypress™ Operators Manual 3-13
The Presets Menu
Storing a User-Defined Preset
To save a user-defined preset, follow these steps:
1. Set each Cypress system function to your preferred setting.
2. Press the SETUP key.
3. Highlight the Presets heading.
4. Click [ADD].
5. Use the Trackball to place the cursor in the Presets text field and type a
unique label or name for the preset. For example, “Deep Venous,” “John’s
Pediatric settings,” “Contrast,” etc.
6. Click [OK].
7. Click [UPDATE].
NOTE A maximum of 16 user-defined presets can be stored.
NOTE Changes can be made to the Presets at any time by changing
the system parameters and then clicking the [
UPDATE] button in
the Setup Presets menu.
Recalling a Saved Preset
To recall a saved Preset, follow these steps:
1. Press the SETUP key.
2. Highlight the Presets heading.
3. Use the pull-down menu to select a Preset option.
4. Click [LOAD]. The Preset is loaded when live imaging begins.
NOTE If a Preset is removed (i.e., deleted) after [LOAD] is clicked and
before the start of live imaging, the factory default Preset will be
loaded.