Transferring Studies
9-8 ACUSON Cypress™ Operators Manual
Changing Server Information
To add and edit server information, use the [ADD] and [EDIT] buttons on the
DICOM menu. For an explanation of the fields in the Cypress Network Identifica-
tion menu, see “Setting an IP Address” on page 9-3.
Adding a Server: Use the [
ADD] button to add a new server. When you click the
button a menu opens. Use the menu to define the new server and Cypress Net-
work Identification information. Each server can have separate Cypress Net-
work Identification information.
Editing a Server: Use the [
EDIT] button to edit settings for an existing server
name. When you click the button a menu opens and you can edit the server and
Cypress Network Identification information.
Removing a Server: Display the server that you want to delete and then click
the [
REMOVE] button.
Renaming a Server: Display the server that you want to rename and then click
the [
RENAME] button. Type the new name in the Network Name field.
DICOM Setup for Network Transfers
If the optional network package is installed, the DICOM Server and Cypress sys-
tem information fields must be completely defined before sending images over a
network. If the network package is not installed, the
DICOM Server and Cypress
system information fields will not be accessible.
To specify your
DICOM settings, follow these steps:
1. Press the SETUP function key.
2. Double-click the DICOM directory listing.
Complete the Server information detailed in steps 3-10:
3. Click the DICOM Server Information subtitle.
4. Click the [ADD] button.
5. Identify the server by typing a user-defined name in the Server Name field.
6. Set the Application Entity. This defines the server application that receives
images from the Cypress system.
7. Set the Server IP Address. This is the IP address of the DICOM server
where images will be sent.
8. Set the Port Number. This is the remote port on the server that is to receive
the connection request. The default setting is 104.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ACUSON Cypress and is the answer not in the manual?

Siemens ACUSON Cypress Specifications

General IconGeneral
BrandSiemens
ModelACUSON Cypress
CategoryMedical Equipment
LanguageEnglish

Related product manuals