EasyManua.ls Logo

Siemens ACUSON Cypress

Siemens ACUSON Cypress
306 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ACUSON Cypress™ Operators Manual 9-7
Exporting Images from the Cypress System
The Cypress DICOM Information fields set the network information for the
Cypress system.
TABLE 42. Cypress DICOM Information Fields
Field Description
Application
Entity
The name the Cypress system presents to the server. For
example, "Cypress_ER1" describes the machine and its
location. This field can include only alphanumeric charac-
ters and the underscore. A maximum of 32 characters are
allowed.
Port Number Identifies the port on the Cypress that receives commands
and data from the server. The number must match the port
number of the server that sends commands and data. The
default setting is 104.
PDU Length
(in bytes)
The Protocol Data Unit Length can be set by changing this
field. The minimum size setting is 4096. The default setting
is also 4096.
Network
Time-out 1
Defines the number of seconds allowed between succes-
sive data blocks when receiving a message. The value is
only used after a part of a message has been received.
The default time-out and minimum value is 5 seconds.
Network
Time-out 2
Defines the number of seconds allowed before a time-out
error is declared on an expected response. The default
time-out and minimum value is 60 seconds.

Table of Contents

Related product manuals