Chapter 5
Maintenance and Repair
Focus 30/35 Service Manual 5 - 14 P/N 79000035_SM, Version 1.5
2 Exchange Receiver
fibre.
A. Loosen the clamp
screw on Receiver
board.
B. Pull out the
Receiver fibre
ferule.
C. Loosen the clamp
screw on Receiver
holder.
D. Pull out the
Receiver fibre
ferule.
E. Unclip the Fiber
bender from the
wall clip.
F. Pull out the
Receiver fibre
complete.
Allen key 0,7mm
5Ncm/0,44lbfin
Ball Allen key
1,3mm
5Ncm/0,44lbfin
Fig. 5-10 Exchange Receiver fibre
3 Remount Receiver
fibre.
Remount the Receiver
fibre in reverse order,
attend the laying see
Figure 5-13 and see Fig.
5-14 on page 5-18
Be careful if you
press in the fibre
bender in to the wall
clip avoiding the
demage of the glas
fibre with the tool
(e.g. tweezer)
4 Adjustment and
calibration
What to do after
replacement?
Item Performance Action Result Notes
A
B
C
D
E

Table of Contents

Other manuals for Spectra Precision Focus 30

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Spectra Precision Focus 30 and is the answer not in the manual?

Spectra Precision Focus 30 Specifications

General IconGeneral
BrandSpectra Precision
ModelFocus 30
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals