Chapter 6
Adjustment and Verification
P/N 79000035_SM, Version 1.5 6 - 153 Focus 30/35 Service Manual
Face 2 panel - Display and Keyboard test
Performance
Fig. 6-116 Control Unit F30 Display, Touch Screen and Keyboard test
9 Exit program Push
Exit button to leave
the program.
Item Performance Action Result Notes
1 Start up instrument /
PASS.
See page 7-3 how to set-
up the instrument.
2 Start Face 2 Display &
Keyboard Test
Select
Display & keyboard
test
from Face 2 panel in
Maintenance & Service, see
Fig. 6-117 on page 6-154 .
3 Start the Display Test A. Select the Display
Test and run Display
Test by pushing the
Start button.
B. Confirm all visible
patterns on the
display with
Yes or
No button.
The Face 2
Display shows
different screen
patterns.
The program
jumps to the
next screen and
shows the
results.