Chapter 6
Adjustment and Verification
Focus 30/35 Service Manual 6 - 156 P/N 79000035_SM, Version 1.5
Final test and report
Start PASS and instrument. Use the tab Final test for all final calibration in sequence. The calibrations
should be done in the recommended sequence as described below.
Tilt sensor offset
HA/VA collimation correction
Trunnion Axis Correction
Tracker collimation
Service Interval
Generate PUK Code for the instrument
Tools and equipment
Tilt sensor offset
Performance
Description Part number
PASS Service software (higher Rev. 2.4 -
Item Performance Action Result Notes
1 Start up instrument /
PASS.
See page 7-3 how to set-
up the instrument.
2 Start Tilt sensor offset A. Select
Tilt sensor
offset from
Instrument setup in
Final Test & Report.
B. Level the instrument
and quit with
OK.
C. Start the tilt offset
determination.
D. Store the correction
values.
E. Measure the tilt
values in both faces
manually.The
residual errors should
less 6cc.
F. Disconnect the
instrument from
PASS or continue
with other final tests.
Tolerance:
X/Y< 30cc
Tolerance:
CorrX/Y< 30cc
The determination
runs automaticely in
both directions.
Attend the messages
in the Program status
box.