Chapter 6
Adjustment and Verification
P/N 79000035_SM, Version 1.5 6 - 103 Focus 30/35 Service Manual
Tracklight
• Tracklight alignment
Tools and equipment
Tracklight alignment
Performance
Description Part number
Prism 62mm SECO Silver coated water sealed <2” 6411-04-BLK
Reticle illumination tool (yellow LED) 79706001
Screw driver Torx T8 Local tool
Item Performance Description Result Notes
1 Start up instrument /
PASS.
See page 7-3 how to set-
up the instrument.
Prepare for Tracklight
alignment.
A. Turn out the
instrument hight
adapter to position 1
of the scale.
B. Place the instrument
on the Collimator
bench.
C. Aim to the cross hair.
The Servo focus
unit have to be
focused to
infinity.
Turn the tribrach
footscrews in a low
hight position.
2 Start the test. Select
Tracklight from
Tracklight in Maintenance
& Service
, see Fig. 6-84 on
page 6-104 .
The Collimator
Monitor
Software starts
automatically.
Refer also to the
Software user guide
chapter in the Trimble
Collimator Bench
manual on:
Collimator Bench
manual
3 Preparation for
Tracklight alignment
with prism or jump to
item
5 using the CM
software.
A. Select
Adjustment
Tracklight check box
and switch on the
diodes.
B. Setup a prism with
diameter 50mm
rectangle aligned to
instrument, see Fig.
6-85 on page 6-106 .
C. Align the correct
prism position using
a piece of paper.
Green and red
diode are
visible.
Distance appr.
0,4m / 15,8in
The prism diameter is
necessary to cover the
offset between
Tracklight and
instrument aperture.