Chapter 6
Adjustment and Verification
P/N 79000035_SM, Version 1.5 6 - 49 Focus 30/35 Service Manual
Preparing instrument and collimator
Performance
Allen key 1,3 mm Local tool
Ball allen key 1,3mm Local tool
Allen key 1,5mm Local tool
Tweezers Local tool
Item Performance Description Result Notes
1 Preparing instrument
and Collimator.
A. Screw out the
instrument hight
adapter to position 3
of the scale, place the
instrument on
Collimator bench.
B. Select
Optical fibre
adjustment Tx/Rx
from EDM DR STD
(3R)
in Maintenance &
Service.
C. Level instrument and
collimator.
D. Switch on Holding
mode.
E. Aim with cross hair
to collimator.
This preparation is
very important prior
the fibre adjustment.