Chapter 6
Adjustment and Verification
Focus 30/35 Service Manual 6 - 168 P/N 79000035_SM, Version 1.5
Service check list
• Focus 30 Check list after service/repair
This chapter includes a check list. The checklist is a guide to be used after service/repair of the
instrument. After the complete check list the instrument should be ready for delivery to the customer.
For information what to do after replacing a service part / module, see page 5-121.
Focus 30 Check list after service/repair
[ ] 1. Check current update on webpage
[ ] 2. Perform Tilt sensor offset
[ ] 3. Perform Servo Focus Unit short range collimation
[ ] 3. Perform HA/VA collimation
[ ] 4. Perform Trunnion axis correction
[ ] 5. Perform Tracker collimation test and function test
[ ] 6. Perform Servo drive test
[ ] 7. Perform an Angle drive test
[ ] 8. Check distances on baselines
[ ] 9. Radio range & short range test - Robotic only.
[ ] 10. Set Service date for next maintenance
[ ] 11. Save constants, error & event log
[ ] 12. Check accessories

Table of Contents

Other manuals for Spectra Precision Focus 30

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Spectra Precision Focus 30 and is the answer not in the manual?

Spectra Precision Focus 30 Specifications

General IconGeneral
BrandSpectra Precision
ModelFocus 30
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals