Chapter 5
Maintenance and Repair
Focus 30/35 Service Manual 5 - 72 P/N 79000035_SM, Version 1.5
Coarse tilt sensor on Slip contacts Hz
• Replacement Coarse tilt sensor
Tools and equipment
Replacement Coarse tilt sensor
Performance
Description Part number
Torx T8 bit Local tool
Torque wrench 120 Ncm T71600280
Item Performance Action Result Notes
1 Remove Side cover
right.
see page 5-27
2 Remove Side cover
left.
see page 5-31
3 Remove Base unit. see page 5-45
4 Remove Slip
contacts Hz with
coarse tilt sensor.
A. Remove three
screws.
B. Remove Slip
contacts Hz.
C. Disconnect Cable
SCH Hz- MCB.
D. Exchange Slip
contacts with
Coarse tilt sensor
and remount it in
reverse order.
35 Ncm/3.1lbfn
Every service part
will be delevered
with a sheet of
paper containing
the Coarse tilt
temperature
calibration values.