Chapter 6
Adjustment and Verification
Focus 30/35 Service Manual 6 - 164 P/N 79000035_SM, Version 1.5
Service Interval
The next Service date can be set in Pass after repair or postponed as customer entry by 3 month on
the Control unit F2 as decribed below.
The
Date and Time in Win CE/ Controle unit F2 overwrites the Date and Time settings of the MCB
board after each starting the software Survey Pro. Please check the settings in Win CE/ Controle unit
F2 under MY Device/Contole panel/Date,Time before you set a new service date.
Performance
1 Start Tracker
collimation
A. Disconnect the
instrument from
PASS, restart the
instrument and run
the test from
SurveyPro as shown
on the instruction.
Follow the
instructions in
the appearing
window.
Execute the operation
in
Survey Pro. The
correction will be
stored as customer
value in the memory
of IM board.
Fig. 6-126 Tracker collimation
Item Performance Action Result Notes
1 Start up instrument /
PASS.
See page 7-3 how to set-
up the instrument.
Item Performance Action Result Notes