Chapter 5
Maintenance and Repair
Focus 30/35 Service Manual 5 - 110 P/N 79000035_SM, Version 1.5
3 Disassembly
Coarse drive lever
arm.
A. Unhook the
extension spring.
B. Unscrew the two
screws of the
divided Coarse
drive lever.
C. Unhook the Coarse
drive lever arm
from the Fine drive
pin of the spindle
and take out the
lever arm.
Put the Instrument
in the Holder
Instrument F30
79297035
Torx T8
60 Ncm/5,31bfn
At this point it is
possible to
exchange the Tilt
sensor without a
complete
disassembling of
the Horizontal
drive.