Chapter 6
Adjustment and Verification
P/N 79000035_SM, Version 1.5 6 - 35 Focus 30/35 Service Manual
correction value and HA/VA collimation have to be done.
Performance
Item Performance Action Result Notes
1 Precondition The reticle is checked see
“Check Reticle
collimation (HA/VA
collimation)” on page 6-
12. and the question
The
vertical collimation is out
of tolerance. Was the
vertical angle system
disassembled or partly
replaced?
was anwered
with
No, .
No correction
values will be
stored in to the
MCB memory.
Condition:
CHz 30” /
0,0083gon
CV > 30” /
0,0083gon
Fig. 6-25 Vertical collimation out of tolerance
Confirm the next message
box with
OK and follow
the steps.