Chapter 6
Adjustment and Verification
P/N 79000035_SM, Version 1.5 6 - 43 Focus 30/35 Service Manual
5 Measure HA/VA
collimation.
Test results:
A. Select Adjust Hz
collimation check
box.
B. Click
Measure Face 2
and aim to the
collimators cross-
hair in face 2.
C. Click
OK to start
measurement.
D. Click
Measure Face 1
and aim to the
collimators cross-
hair in face1.
E. Click
OK to start
measurement.
F. Check test results and
confirm message
box:
Case 1:
Collimation error
CHz 83cc and CV
< 83cc. Confirm the
message box with
Yes and go to item 6
to adjust the reticle in
horizontal direction.
Case 2:
Collimation error
CHz 83cc and CV
> 83cc. Confirm the
message box with
Yes and go to item 6
to adjust the reticle in
horizontal direction.
Repeat the
measurment with
answering
No.
The instrument
turns to face 2
automatically.
The instrument
turns to face 1
automatically.
The collimation
error CHz/CV
and the
adjustment
position AP Hz
will be shown.
Message box
appears to store
the correction
values if CHz
<30cc and CV
<83cc.
Message box
appears to store
the CV
correction
values.
The measurement can
be repeated and the
correction value
stored after each
measurement of face
2 / face 1.
The correction value
CV should be stored
only, if the
adjustment with the
Front lens mirror was
done successfully!