Chapter 6
Adjustment and Verification
P/N 79000035_SM, Version 1.5 6 - 65 Focus 30/35 Service Manual
2 Mount tools. Remove the Cover front
lens and mount Adapter
Focus 30 and Scattered
light device tool on the
Front lens.
Keep the inner side of
the Scattered light
device clean, because
fingerprints and
grease can affect the
signal and thereby the
result.
Fig. 6-51 Scattered light device mounted on instrument
3 Check Reference fibre
signal.
A. Select
Reference fibre
adjustment
from EDM
DR STD (3R) in
Maintenance &
Service
.
B. Click
Holding Mode
ON.
C. Click Fibre
adjustment
Start.
D. Check the Signal
(ext./int).
The extern/
intern signal
values will be
displayed.
Item Performance Description Result Notes
Scattered
light device
Adapter