Chapter 6
Adjustment and Verification
P/N 79000035_SM, Version 1.5 6 - 93 Focus 30/35 Service Manual
1 Start up instrument /
PASS.
A. Mount the instrument
on the Collimator
bench and See page
7-3 how to set-up the
instrument.
B. Remove Servo focus
unit cover See
“Telescope covers”
on page 5-2.
C. Select
Tracker
receiver alignment
from Tracker
adjustment
in
Maintenance &
Service.
D. Turn instrument in
face II and to the
collimator.
E. Click Holding mode
ON
Screw out the
instrument hight
adapter to position 3
of the scale.
2 Start Tracker receiver
camera.
A. Select
Switch Tracker
receiver ON/OFF
checkbox.
B. Click Tracker
receiver
ON.
C. Click Find Optimal
Camera shutter
Start
and the software is
calculating the
correct exposure time
visible on the moving
slider. The shutter
position can be
repeated pushing
Start again.
D. The Collimator
pattern can be
magnified with
selecting the
checkbox
Zoom
Image.
The software
shows on
Result/
Staus
the
collimators
crosshair pattern
via Tracker
receiver camera.
The collimator
pattern should
be correct
visible.
The refresh rate
causes a time delay of
any adjustment
action.
If the pattern is to
dark or bright adjust
the setting on your PC
monitor.
Item Performance Action Result Notes