SVAN 948 USER MANUAL _
5 - 32
In each available position any change is performed by means of the <
>, <
> push-buttons. In
order to confirm the selection the <ENTER> push-button has to be pressed. After this confirmation the
opened window or list is closed. In order to ignore any changes made in the opened window or list the
user has to press the <ESC> push-button.
Selection of the parameters of vibration dose - MEASURE DOSE
In the EXPOSURE TIME position (path: MENU / SETUP / HAV/WBV DOSE / MEASURE DOSE) there
are two possibilities: the user can switch on (ENABLED) or switch off (DISABLED) the measurement of
vibration dose. If the user selects ENABLED in MEASURE DOSE, he should also select the correct filter
in INPUT/ CHANNELS SETUP/ CHANNEL x. If it is chosen an incorrect filter, it appears on the display
the message: INCORRECT SETUP, VIB. DOSE IS OFF and measurements of dose are not done.
Otherwise the measurement is continued.
The view of the display informing about wrong selected setting
The measure (MEASURE DOSE) can be switched on or switched off using the <
>, <
> push-
buttons. In order to confirm the setting the <ENTER> push-button has to be pressed. Such pressing
closes the sub-list. After pressing the <ESC> push-button the sub-list is also closed but all changes which
were made are ignored.
The view of the displays in the MEASURE DOSE window, the selection of setting
Selection of the parameters of vibration dose - EXPOSURE TIME
In the EXPOSURE TIME position (path: MENU / SETUP / HAV/WBV DOSE / EXPOSURE TIME) the
user can set the exposure time of the vibration signal. It is possible to set maximum 24 hours with 1 s step
pressing the <
>, <
> push-buttons or with 30 s step pressing the <SHIFT> with <
>, <
> push-
buttons. In order to confirm the setting the <ENTER> push-button has to be pressed. Such pressing
closes the sub-list. After pressing the <ESC> push-button the sub-list is also closed but all changes which
were made are ignored.
The view of the displays during the setting of the EXPOSURE TIME
(path: MENU/ SETUP/ HAV/WBV DOSE/ EXPOSURE TIME); exposure time expressed in hours