SVAN 948 USER MANUAL .
3 - 8
o TOTAL 3: (sub-list) enables one to select the weighted filter to be used for the
calculation of the TOTAL value in the third profile:
FILTER: (position); available values of the weighted filters:
in the case of sound measurements: LIN, SUSR1, SUSR2, SUSR3 or any
other sent to the unit by means of the interface
in the case of vibration measurements: CH, VUSR1, VUSR2, VUSR3 or
any other sent to the unit by means of the interface
TYPE: (position available only for vibration measurements); available values
if VUSR1, VUSR2 or VUSR3 was selected in the previous position: ACC,
VEL and DIL; if the CH filter was selected (the filter which was set in the 3
rd
channel) this position is not displayed
CAL. F.: (position available only for vibration measurements); accessible if
VUSR1, VUSR2 or VUSR3 was selected in the FILTER position; if the CH
filter was selected (the filter which was set in the 3
rd
channel) this position is
not displayed; available values from -60.0dB to 60.0dB with 0.1dB or 1 dB
step
SPECTRUM TYPE: (position); available values of this position:
ACCELERATION, VELOCITY and DISPLACEMENT - in the case of
vibration measurements and
LEQ, RMS - in the case of sound measurements
BUFFER VIEW (sub-list)
§ NO.: (position); available number of the files in the instrument’s buffer containing the results
of measurements
§ RECS (position); it informs the user how many records with the measurement results contains
the selected file from the instrument’s buffer)
§ FREE (position); it informs the user about the size of the available memory in the instrument’s
buffer
BATTERY (window); it informs the user about the state of the internal battery of the instrument
CONTRAST (position); it enables the user to select one from ten possibilities of the contrast of the
instrument’s display
BACKLIGHT (sub-list)
§ TIMEOUT (position), available values [
] or [ ], if [
] is chosen it will cause the self made
backlight switching off in the case when the keyboard is not used during the last 30 seconds.
If it happened the first pressing of any push-button switches the backlight on
§ BRIGHTNESS (position); it enables the user to select one from four possibilities of the
brightness of the instrument’s backlight; the backlight is used both for the display and for the
keyboard
UNIT LABEL (window); it informs the user about the serial number of the unit, the internal software
version, the internal memory size and the standards which the instrument fulfils

Table of Contents