7.11.6.3 Zx Multi-frequency Setup
If the user needs to use different frequencies to test different Zx pins of the transformer, in the
transformer Zx limit setting table, touch or move the cursor to the frequency area of Zx, after
inputting the frequency in this area, the frequency here will be used automatically during the test. If
the parameter is not set, the frequency of Zx in the test condition table will be used.
7.11.6.4 Zx Multi-level Setup
If the user needs to use different levels to test different Zx pins of the transformer, in the transformer
Zx limit setting table, touch or move the cursor to the level area of Zx, after inputting the level in
this area, it will be used automatically during the test. If the parameter is not set, the level of Zx in
the test condition table will be used.