161
:CORR:OPEN:STAT <0|1|ON|OFF>
Parameter:
0|OFF ---- OFF
1|ON ---- ON
Example:
:CORR:OPEN:STAT 0 ----Set the open circuit clearing state to ON
:CORR:OPEN:STAT 1 ----Set the open circuit clearing state to OFF
:CORR:OPEN:STAT? ----Return to open circuit clearing switch status
9.2.17.2 SHORT
Description: Execute preset test point short circuit clearing
Command systax:
:CORR:SHOR [ACK]
Parameter:
Example:
:CORR:SHOR ---- Perform short-circuit clearing operation on preset point, no return
:CORR:SHOR ACK---- Perform short-circuit clearing operation on preset point,
Return to
Description: Set the state of the short-circuit correction switch
Command systax:
:CORR:SHOR:STAT?
:CORR:SHOR:STAT <0|1|ON|OFF>
Parameter:
0|OFF ---- OFF
1|ON ---- ON
Example:
:CORR:SHOR:STAT 0 ----Set the short-circuit clearing state to ON
:CORR:SHOR:STAT 1 ----Set the short-circuit clearing state to OFF
:CORR:SHOR:STAT? ---- Return to short circuit clearing switch status
9.2.17.3 LOAD
Description: Set the load correction switch state
Command systax:
:CORR:LOAD:STAT?
:CORR:LOAD:STAT <0|1|ON|OFF>
Parameter:
0|OFF ---- OFF
1|ON ---- ON
Example:
:CORR:LOAD:STAT 0 ----Set the load correction status to ON
:CORR:LOAD:STAT 1 ----Set the load correction status to OFF
:CORR:LOAD:STAT? ----Return to load correction switch status