18.THE LIST OF INITIAL SETTING VALUE
177
18-3. Setting Mode 2
18-4. Setting Mode 3
18-5. Setting Mode 4
*D/A Zero Output Weight and D/A Full Scale cannot be set larger value than the value of Capacity.
The value of Capacity automatically reaches a set value.
→→→
F
CNG/ENT CNG/ENT
2
LOWER
Setting item
Initial
value
LOCK 1
(SRAM)
LOCK 2
(NOV.RAM)
LOCK SW
(NOV.RAM)
Display
1 Weighing Function 1 0000
◎
2 Weighing Function 2 01000
◎
3 Weighing Function 3 00141
◎
4 Sequence Mode 0000
◎
5 Function Key Invalid 1111
◎
6 Digital Filter 64
◎
7 Motion Detection 1.5-05
◎
8 Zero Tracking Period 0.0
◎
9 Zero Tracking Range 0000
◎
→→→
F
CNG/ENT CNG/ENT
3
NEAR Z
.
Setting item
Initial
value
LOCK 1
(SRAM)
LOCK 2
(NOV.RAM)
LOCK SW
(NOV.RAM)
Display
1 Balance Weight Value 100.00
◎◎
2 Capacity 100.00
◎◎
3 Min. Scale Division 0.01
◎◎
4 Net Over 999.99
◎◎
5 Gross Over 999.99
◎◎
6 DZ Regulation Value 02.00
◎◎
7 Function Selection 13213
◎
8
Gravitational Acceleration
(area number input)
09
◎
9
Gravitational Acceleration
(acceleration input)
9.798
◎
Setting item
Initial
value
LOCK 1
(SRAM)
LOCK 2
(NOV.RAM)
LOCK SW
(NOV.RAM)
Display
1 D/A Output Mode 00
◎
2 D/A Zero Output Weight 00000
◎
3 D/A Full Scale 10000
◎
4 RS-485 I/F 30101
◎
5 ID Number 0000
◎
6 RS-232C I/F 30101
◎
7 External Function Selection 00000
◎
8 Setting Value LOCK 00
← LOCK 1 and LOCK 2 are set up (it memorizes to NOV.RAM)
9
Restriction on the Tare
Subtraction Function
0000
◎