5.FUNCTION SETTINGS
63
5-11. One-touch Tare Subtraction
Tare is subtracted and Net weight is zeroed by pressing key.
However, the operation is performed only when “ ” is on depending on the setting of
Restriction on the Tare Subtraction Function in setting mode 4. The range of Tare subtraction can be
selected from all range or 0 < Tare ≦ Capacity.
1) Press .
2) “ ” lights.
- Tare subtraction by external signal
By short-circuiting A4pin (Tare subtraction ON) and COM on the control connector, Net weight is
immediately zeroed, and “ ” lights.
* If Net weight is not zeroed by performing the operation of Tare subtraction, the following causes
are presumable.
Cause Measure
The indicated value is Gross weight.
Press the key to change the
indicated value into Net weight.
(If “ ” is on, the indicated value is Net
weight.)
The indicated value is not stable.
(By setting)
Perform the operation of Tare subtraction
when “ ” is ON.
The indicated value is out of the
range of Tare subtraction.
(By setting)
Perform the operation of Tare subtraction in
the range of Tare subtraction.
OFF
ON
Pulse width 50msec or more