EasyManuals Logo

Unipulse F701-C User Manual

Unipulse F701-C
189 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #113 background imageLoading...
Page #113 background image
7.STATISTICAL AND ACCUMULATION FUNCTION SETTING AND OPERATION
102
7. STATISTICAL AND ACCUMULATION FUNCTION
SETTING AND OPERATION
Weight values are accumulated by this function. Based on the accumulated weight data, statistical
data, such as the maximum, minimum, and average, are calculated.
Weight values can be accumulated by the auto accumulation command or external input/output
signal, through RS-232C communication, RS-485 communication (option), CC-Link (option).
Up to 10000 accumulations can be made. For 10001 or more accumulations, the operation
“Accumulation clear” must be performed to clear the accumulated and statistical data.
While the number of accumulations is 10000, an accumulation error results. The accumulation error
can be outputted from B11 pin out of the external input/output signals (by setting).
Even while the accumulation error is given, other weighing operations and the print command (SI/
F), etc., are not affected.
Operations other than those relating to statistics and accumulations are performed normally.
The accumulation value and statistical data can be referred to in the subdisplay area or in the setting
mode 8.
(For details, please refer to "1-1.Front Panel"Subdisplay on P.1, "3-2-7.Setting Mode 8" on P.38.)
The accumulated weight could be Gross or Net weight, which is
selected in Over / Under Comparison of the Comparison Mode.

Table of Contents

Other manuals for Unipulse F701-C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Unipulse F701-C and is the answer not in the manual?

Unipulse F701-C Specifications

General IconGeneral
BrandUnipulse
ModelF701-C
CategoryAccessories
LanguageEnglish

Related product manuals