4.CALIBRATION
50
4-4-8. 1/4 Scale Division
It divides the minimum scale division into four (4) parts. The “ ” (center zero) lamp turns on
when the weight is between +1/4 division and -1/4 division.
1/4 scale division selects ON/OFF.
1) Select setting mode 3.
2) Select 1/4 scale division display ON/OFF. (7-Function Selection)
→→→
F
CNG/ENT CNG/ENT
3
NEAR Z
.
→→ →
1/4 scale division display
1 : ON
0 : OFF
ZERO
7
OVER
CNG/ENTCNG/ENT
ZERO ZERO
nn+1
1/4 scale
Min. Scale Division
ON
“ ”
(1scale divisio)
ZERO
ON
“ ”
ZERO
When the 1/4 scale division setting is OFF, “ “ lamp only works at the zero
point.