EasyManuals Logo

Unipulse F701-C User Manual

Unipulse F701-C
189 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #23 background imageLoading...
Page #23 background image
2.CONNECTION
12
2-2. Connection of the Power Input Terminals
Connect AC power cord. The input voltage is 100V ~ 240V AC.
The frequency is 50/60Hz.
1) Make sure that no power is applied.
2) Remove the terminal block cover.
3) Remove the two screws(M3) of the terminal block, align the crimp contacts with the
screw holes, and then tighten the screws.
4) Mount the terminal block cover.
5) Remove the screws(M4) of the protective ground, align the crimp contacts with the
screw holes, and then tighten the screws.
2-3. Connection of the Guard Ground
The grounding terminal is for prevention of electric shocks.
Use an approx. 0.75mm
2
thick wire, and be sure to ground.
block cover
Within 6mm
AC power cord
Frame
ground
- Connect with no power applied because it may cause an electric shock.
- The attached AC cable is designed for domestic use in Japan, and its rating is 125V AC,
7A. For use at voltages exceeding the rating and for overseas use, have a separate AC
cable prepared.
- Since the F701-C has no power switch, install a breaker.
- Be sure to ground the protective ground terminal to prevent electric shocks.
Do not use other screws than that attached to the main body.
CAUTION

Table of Contents

Other manuals for Unipulse F701-C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Unipulse F701-C and is the answer not in the manual?

Unipulse F701-C Specifications

General IconGeneral
BrandUnipulse
ModelF701-C
CategoryAccessories
LanguageEnglish

Related product manuals