4.CALIBRATION
52
- “ ”
Inditial dead load is negative.
Check that load cell is mounted in the correct direction; check that load is being applied to
the load cell in the correct direction; check that the +SIG and -SIG lines are propely
connected.
If still displayed, connect a resistor between -EXC and -SIG load cell
connections. This should shift the zero point. Do Zero Calibration again.
- This table is for a 350 Ω load cell.
- The temperature coefficient of the connected resistor directly influences the accuracy
of the controller. Use a resistance having a temperature coefficient of 50ppm/ ℃ or
more (recommended value of about 5ppm/ ℃ ).