EXAMPLE: SET UP A 5,000 LB. CAPACITY SCALE WITH A
GRADUATION SIZE OF 0.5 LB. ON CHANNEL 1.
1. Enter Calibration Mode for Channel 1.
2. Select the weighing units for Set-Up by pressing 7 followed by the
[PRINT/SELECT] key. Either the lb or kg LED indicator LED will illuminate.
3. To select Decimal Point Position, press 2 on the keypad followed by
[PRINT/SELECT]. The display will show the current position (default: 0).
4. For 1 decimal place, press 1 on the keypad followed by the [PRINT/SELECT]
key.
5. To select Graduation size, press 3 on the keypad followed by [PRINT/SELECT].
The display will show the current Graduation size (default: 1)
6. For a grad size of half a pound, press 5 on the keypad followed by
[PRINT/SELECT].
7. To select Scale Capacity, press 4 on the keypad followed by [PRINT/SELECT].
The display will show the current Scale Capacity.
8. For a Scale Capacity of 5,000 lb., enter 5000 on the keypad followed by the
[PRINT/SELECT] key.
The Scale is now ready for calibration and/or more advanced Set-Up.