57
TECHNICAL MANUAL
260P
fac Re5et
Pa55
fac init
1P
cal CLEAR Pa55
Cal 1
init
RELOAD FACTORY CALIBRATION SETTINGS - PARAMETER 1
Description: Clears all previous Calibration values for the selected Scale
Channel. The User will be prompted for the Password.
1. Enter Parameter 260.
2. Enter the Password.
3. The M2000 will restart.
The following Parameters will be reset:
Span & Span Table entries (13, 110) Standard Scale Set-Up (2, 3, 4)
Motion (5, 24) Units (6, 7)
Zero & Zero Tracking (12, 21, 22, 23) Scale Filtering (41, 42, 43, 44)
COMPLETE FACTORY INITIALIZATION - PARAMETER 260
Description: Completely resets all M2000 parameters to factory default
values. All previous Calibration information will be lost.
1. Enter Parameter 260.
2. Enter the Password.
3. The M2000 will remain in Calibration Mode.
ALL PREVIOUSLY SAVED SETTINGS & CALIBRATION IN THE
M2000 WILL BE LOST! MAKE SURE ANY IMPORTANT
SETTINGS ARE RECORDED BEFORE ACTIVATING THIS
PARAMETER!