SMARTWIRE SET-UP PARAMETERS
ENABLING SMARTWIRE - PARAMETER 59
Description: Opens or closes the SMARTWIRE port.
PARAMETER VALUE ACTION
0 (default) Disable SMARTWIRE port
1 Enable SMARTWIRE port
SMARTWIRE FAILURE - ERROR 185
If there is a component failure, or the SMARTWIRE connection is lost due to
breakage of the communications wire, āERR185ā will be displayed. This error
message will continuously appear until a new link is established.
To regain control of the indicator, you may have to disable the SMARTWIRE
port. This can be done in Calibration or Normal Weighing Mode.
1. Press [CLEAR] to temporarily get rid of the Error message.
2. Enter Parameter 59.
3. Enter [0] followed by the [PRINT/SELECT] key to disable.
SMARTWIRE must be enabled to communicate with any
connected device.
IF NO DEVICE IS PLUGGED INTO THE ENABLED
SMARTWIRE PORT, A COMMUNICATIONS TIMEOUT WILL
OCCUR PRODUCING AN ERROR (ERR 185).