TECHNICAL MANUAL
MAIN FILTER (DIGITAL AVERAGING) SETTING - PARAMETER 41
Description: Adjusts the level of filtering (averaging) performed to get the
final weight.
Increasing the filter value increases the number of weight samples
used for the average. At a time of sustained weight, the display
value will change very little.
Changing the filter value alters the settling time for the final weight.
Accepted Values: 1 to 255
PARAMETER VALUE RESULT OF FILTERING
1
Very little Main Filter. M2000 will be extremely fast,
but will likely not settle.
64 (default)
Mid-range. Suitable for most applications.
255
Heavy Main Filter. M2000 will be very slow in any
small weight change.
FASTSTEP FILTER THRESHOLD - PARAMETER 42
Description: Controls the how the M2000 switches from the Main Filter to
the Faststep Filter.
The threshold value is the number of graduations the weight
on the scale must change before the Faststep Filter kicks in.
Default Value: 8 graduations.
IF THE FASTSTEP FILTER THRESHOLD VALUE IS TOO LOW,
WEIGHT READINGS WILL BE UNSTABLE REGARDLESS OF
THE MAIN FILTER SETTING!